We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVE: To compare the defibrillation efficacy of a novel lead system placed in the middle cardiac vein with a conventional non-thoracotomy lead system. METHODS: In eight pigs (weighing 35-71 kg), an electrode was advanced transvenously to the right ventricular apex (RV), with the proximal electrode in the superior caval vein (SCV). Middle cardiac vein (MCV) angiography was used to delineate the anatomy before a three electrode system (length 2 x 25 mm + 1 x 50 mm) was positioned in the vein. An active housing (AH) electrode was implanted in the left pectoral region. Ventricular fibrillation was induced and biphasic shocks were delivered by an external defibrillator. The defibrillation threshold was measured and the electrode configurations randomised to: RV-->AH, RV+MCV-->AH, MCV-->AH, and RV-->SCV+AH. RESULTS: For these configurations, mean (SD) defibrillation thresholds were 27.3 (9.6) J, 11.9 (2.9) J, 15.2 (4.3) J, and 21.8 (9.3) J, respectively. Both electrode configurations incorporating the MCV had defibrillation thresholds that were significantly less than those observed with the RV-->AH (p < 0.001) and RV-->SCV+AH (p < 0.05) configurations. Necropsy dissection showed that the MCV drained into the coronary sinus at a location close to its orifice (mean distance = 2.7 (2.2) mm). The MCV bifurcated into two main branches that drained the right and left ventricles, the left branch being the dominant vessel in the majority (6/7) of cases. CONCLUSIONS: Placement of specialised defibrillation electrodes within the middle cardiac vein provides more effective defibrillation than a conventional tight ventricular lead.

Original publication

DOI

10.1136/heart.84.4.425

Type

Journal article

Journal

Heart

Publication Date

10/2000

Volume

84

Pages

425 - 430

Keywords

Analysis of Variance, Animals, Coronary Vessels, Corrosion Casting, Defibrillators, Implantable, Electrodes, Equipment Design, Evaluation Studies as Topic, Female, Swine, Ventricular Fibrillation