We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVES: We used optical coherence tomography, which has a resolution of <20 microm, to analyze thin layers of neointima in rapamycin-eluting coronary stents. BACKGROUND: Lack of neointimal coverage has been implicated in the pathogenesis of drug-eluting coronary stent thrombosis. Angiography and intracoronary ultrasound lack the resolution to examine this. METHODS: We conducted a randomized trial in patients receiving polymer-coated rapamycin-eluting stents (Cypher, Cordis, Johnson & Johnson, Miami, Florida) and nonpolymer rapamycin-eluting stents (Yukon, Translumina, Hechingen, Germany) to examine neointimal thickness, stent strut coverage, and protrusion at 90 days. Twenty-four patients (n = 12 for each group) underwent stent deployment and invasive follow-up at 90 days with optical coherence tomography. The primary end point was binary stent strut coverage. Coprimary end points were neointimal thickness and stent strut luminal protrusion. RESULTS: No patient had angiographic restenosis. For polymer-coated and nonpolymer rapamycin-eluting stents, respectively, mean (SD), neointimal thickness was 77.2 (25.6) microm versus 191.2 (86.7) mum (p < 0.001). Binary stent strut coverage was 88.3% (11.8) versus 97.2% (6.1) (p = 0.030). Binary stent strut protrusion was 26.5% (17.5) versus 4.8% (8.6) (p = 0.001). CONCLUSIONS: Mean neointimal thickness for the polymer-coated rapamycin-eluting stent was significantly less than the nonpolymer rapamycin-eluting stent but as a result coverage was not homogenous, with >10% of struts being uncovered. High-resolution imaging allowed development of the concept of the protrusion index, and >25% of struts protruded into the vessel lumen with the polymer-coated rapamycin-eluting stent compared with <5% with the nonpolymer rapamycin-eluting stent. These findings may have important implications for the risk of stent thrombosis and, therefore, future stent design. (An optical coherence tomography study to determine stent coverage in polymer coated versus bare metal rapamycin eluting stents. ORCA 1, from the Optimal Revascularization of the Coronary Arteries group; ISRCTN42475919).

Original publication

DOI

10.1016/j.jcin.2009.01.010

Type

Journal article

Journal

JACC Cardiovasc Interv

Publication Date

05/2009

Volume

2

Pages

437 - 444

Keywords

Confidence Intervals, Coronary Artery Disease, Coronary Restenosis, Coronary Vessels, Drug-Eluting Stents, Endothelium, Vascular, Female, Humans, Immunosuppressive Agents, Male, Middle Aged, Polymers, Sirolimus, Tomography, Optical Coherence