We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVES: The aims of this study were to detect the acetylcarnitine resonance line at 2.13 ppm in the human vastus lateralis and soleus muscles, assess T1 and T2 relaxation times, and investigate the diurnal and exercise-related changes in absolute concentration noninvasively, using proton magnetic resonance spectroscopy at 7 T. MATERIALS AND METHODS: All measurements were performed on a 7 T whole-body Magnetom MR system with a 28-channel knee coil. Five healthy, moderately trained volunteers participated in the assessment of the detectability, repeatability, and relaxation times of acetylcarnitine. For the evaluation of the effect of training status, another 5 healthy, normally active volunteers were examined. In addition, normally active volunteers underwent a day-long protocol to estimate diurnal changes and response to the exercise. RESULTS: Using a long echo time of 350 milliseconds, we were able to detect the acetylcarnitine resonance line at 2.13 ppm in both muscle groups without significant lipid contamination. The T1 of acetylcarnitine in the vastus lateralis muscle was found to be 1807.2 ± 513.1 milliseconds and T2 was found to be 129.9 ± 44.9 milliseconds. Concentrations of acetylcarnitine from the vastus lateralis muscle in moderately trained volunteers were higher than concentrations from normally active volunteers. Acetylcarnitine concentrations changed during the day, tending to be higher in the morning after an overnight fast than after lunch. After 10 minutes of high-intensity exercise, the concentration significantly increased, and 15 minutes after cessation of exercise, a decrease could be observed. CONCLUSIONS: Our results demonstrate an effective detection of acetylcarnitine using a long TE of 350 milliseconds at 7 T in the vastus lateralis and soleus muscles with high repeatability and reliability on a 7 T scanner. Our data emphasize the need for strict standardization, physical activity, and dietary conditions for the measurement of the acetylcarnitine.

Original publication

DOI

10.1097/RLI.0000000000000355

Type

Journal article

Journal

Invest Radiol

Publication Date

07/2017

Volume

52

Pages

412 - 418

Keywords

Acetylcarnitine, Adult, Exercise, Female, Humans, Male, Muscle, Skeletal, Proton Magnetic Resonance Spectroscopy, Quadriceps Muscle, Reference Values, Reproducibility of Results, Time