We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

A method is reported for the measurement of quercetin in human plasma using reversed-phase high-performance liquid chromatography (HPLC). Quercetin and kaempferol (as internal standard) were spiked into plasma samples and extracted using C18 Sep-Pak Light cartridges (efficiency > 85%). Flavonoids were eluted with aqueous acetone (50% v/v, pH 3.5), dried down and redissolved in aqueous acetone (45% v/v, pH 3.5). The increased osmolarity promoted a phase separation and the water-saturated acetone layer, containing the flavonoids, was analysed by HPLC with aqueous acetone mobile phase (45% v/v acetone in 250 mM sodium dihydrogen sulphate. The mixture was adjusted to pH 3.5 with phosphoric acid and used at a flow-rate of 1.0 ml/min) and mu Bondapak C18 column (150 x 3.9 mm I.D., 10 microns particle size). The detection limit (A375 nm) for quercetin in plasma was 0.1 microgram/ml (300 nM). The method also detects metabolites of quercetin, although these are not yet identified.

Original publication

DOI

10.1016/0378-4347(94)00549-k

Type

Journal article

Journal

J Chromatogr B Biomed Appl

Publication Date

07/04/1995

Volume

666

Pages

149 - 155

Keywords

Chromatography, High Pressure Liquid, Humans, Quercetin, Reference Standards, Reproducibility of Results, Spectrometry, Fluorescence